Performance Tests
4- 50
TDS6000 Series Service Manual
Equipment
required
One leveled sine-wave generator (Item 12)
One 2X attenuator (Item 26)
Four TCA-BNC adapters (Item 19)
Three 50
Ω
terminators (Item 3)
One 50
Ω
, precision coaxial cable (Item 4)
Prerequisites
See page 4--17
TDS6000 Series
50
Ω
Coaxial cable
Output
Leveled
Sine Wave
Generator
2X Attenuator
50
Ω
Terminators
Figure 4- 18: Initial test hookup
1.
Install the test hookup and preset the instrument controls:
a.
Initialize the oscilloscope:
Press the
DEFAULT SETUP
button.
b.
Modify the initialized control settings:
H
Turn on all vertical channels (press the Vertical button of any off
channels: CH 1, CH 2, CH 3, and CH 4).
H
Set the Horizontal SCALE to
1.25 ns
.
H
Set the Vertical SCALE of CH 1, CH 2, CH 3, and CH 4 to
100 mV
.
H
From the tool bar, touch
Horiz
and select the
Acquisition
tab.
H
Touch
Average
and set the number of averages to
16
.
H
Touch the
Equivalent Time ET
button.
H
From the button bar, touch
MEAS
.
H
Touch the Source Channel
1
button.
Check Channel Isolation
(Crosstalk)
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...