Performance Tests
TDS6000 Series Service Manual
4- 53
H
Set the Horizontal SCALE to 2 to 5 cycles of the signal.
H
Set the generator output at a sine wave of
4 GHz
bandwidth to check
your oscilloscope. Set the test signal amplitude for about five
divisions on screen. Now fine adjust the generator output until the
CH 1 Amplitude
readout indicates the amplitude is 500 mV.
Readout may fluctuate around 500 mV.
H
Remove the 2X attenuator.
5.
Confirm the input channels are within limits for channel isolation:
a.
Check — Amplitude of each trace other than CH 1 is 0.5 division or less
(discount trace width). Enter the largest amplitude on the test record.
b.
Move the signal to the CH 2 input connector, change the Trigger
SOURCE to CH 2, and move the 50
Ω
terminator to the CH 1 input.
c.
Check — Amplitude of each trace other than CH 2 is 0.5 division or less
(discount trace width). Enter the largest amplitude on the test record.
d.
Move the signal to the CH 3 input connector, change the Trigger
SOURCE to CH 3, and move the 50
Ω
terminator to the CH 2 input.
e.
Check — Amplitude of each trace other than CH 3 is 0.5 division or less
(discount trace width). Enter the largest amplitude on the test record.
f.
Move the signal to the CH 4 input connector, change the Trigger
SOURCE to CH 4, and move the 50
Ω
terminator to the CH 3 input.
g.
Check — Amplitude of each trace other than CH 4 is 0.5 division or less
(discount trace width). Enter the largest amplitude on the test record.
h.
Select an unchecked Vertical SCALE:
H
Connect the 2X attenuator to CH 1.
H
Move the signal to CH 1, set the Trigger SOURCE to CH 1, and
move the 50
Ω
terminator to the CH 4 input.
H
Press
PUSH TO SET 50%
.
H
Set the Vertical SCALE of CH 1, CH 2, CH 3, and CH 4 to
50 mV
H
Set the test signal amplitude for about five divisions on screen. Now
fine adjust the generator output until the
CH 1 Amplitude
readout
indicates the amplitude is 250 mV. Readout may fluctuate around
250 mV.
H
Remove the 2X attenuator.
H
Repeat steps a through g.
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...