Specifications
1- 20
TDS6000 Series Service Manual
Table 1- 12: Certifications and compliances (Cont.)
Category
Standards or description
EC Declaration of Conformity --
Low Voltage
Compliance was demonstrated to the following specification as listed in the Official Journal of the
European Union:
Low Voltage Directive 73/23/EEC, amended by 93/68/EEC
EN 61010-1/A2:1995
Safety requirements for electrical equipment for measurement
control and laboratory use.
U.S. Nationally Recognized
Testing Laboratory Listing
UL3111-1, First Edition
Standard for electrical measuring and test equipment.
Canadian Certification
CAN/CSA C22.2,
Safety requirements for electrical equipment for measurement,
No. 1010.1-92
control, and laboratory use.
Additional Compliance
IEC61010-1
Safety requirements for electrical equipment for measurement,
control, and laboratory use.
Installation (Overvoltage)
Category
Terminals on this product may have different installation (overvoltage) category designations. The
installation categories are:
CAT III
Distribution-level mains (usually permanently connected). Equipment at this level is
typically in a fixed industrial location.
CAT II
Local-level mains (wall sockets). Equipment at this level includes appliances, portable
tools, and similar products. Equipment is usually cord-connected.
CAT I
Secondary (signal level) or battery operated circuits of electronic equipment.
Pollution Degree
A measure of the contaminates that could occur in the environment around and within a product.
Typically the internal environment inside a product is considered to be the same as the external.
Products should be used only in the environment for which they are rated.
Pollution Degree 2
Normally only dry, nonconductive pollution occurs. Occasionally a
temporary conductivity that is caused by condensation must be
expected. This location is a typical office/home environment.
Temporary condensation occurs only when the product is out of
service.
Safety Certification Compliance
Equipment Type
Test and measuring
Safety Class
Class 1 (as defined in IEC 61010-1, Annex H) -- grounded product
Pollution Degree
Pollution Degree 2 (as defined in IEC 61010-1). Note: Rated for indoor use only.
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...