Performance Tests
TDS6000 Series Service Manual
4- 67
H
Press the Trigger
MODE
button to toggle it to
Normal
.
H
From the tool bar, touch
Horiz
and select the
Acquisition
tab.
H
Touch
Average
and set the number of averages to
16
.
H
Touch the
Equivalent Time ET
button.
c.
Hook up the test-signal source:
H
Connect the signal output of the generator to a BNC T connector.
Connect one output of the T connector to
CH 1
through a 50
Ω
precision coaxial cable and an adapter. Connect the other output of
the T connector to a 50
Ω
terminator. Connect the terminator to the
AUX INPUT
. See Figure 4--26.
TDS6000 Series
Sine Wave
Generator
To AUX IN
50
Ω
Terminator
Figure 4- 26: Initial test hookup
2.
Confirm the trigger system is within sensitivity limits (50 MHz):
a.
Display the test signal:
H
Set the generator frequency to 50 MHz.
H
From the button bar, touch
MEAS
.
H
Touch Setup
Ref Levs
; then touch the
Min-Max
button.
H
Touch the
Setup
button and select the
Ampl
tab; then touch the
Amplitude
button.
H
Touch
Close
.
H
Press
PUSH TO SET 50%
.
Summary of Contents for TDS6404
Page 4: ......
Page 14: ...Table of Contents x TDS6000 Series Service Manual...
Page 18: ...Service Safety Summary xiv TDS6000 Series Service Manual...
Page 52: ...Operating Information 2 12 TDS6000 Series Service Manual...
Page 56: ...Theory of Operation 3 4 TDS6000 Series Service Manual...
Page 60: ...Performance Verification 4 4 TDS6000 Series Service Manual...
Page 84: ...Performance Tests 4 28 TDS6000 Series Service Manual...
Page 150: ...Adjustment Procedures 5 2 TDS6000 Series Service Manual...
Page 156: ...Maintenance 6 6 TDS6000 Series Service Manual...
Page 170: ...Removal and Installation Procedures 6 20 TDS6000 Series Service Manual...
Page 228: ...Repackaging Instructions 6 78 TDS6000 Series Service Manual...
Page 232: ...Options 7 4 TDS6000 Series Service Manual...
Page 234: ...Electrical Parts List 8 2 TDS6000 Series Service Manual...
Page 252: ...Mechanical Parts List 10 16 TDS6000 Series Service Manual...