![Tektronix SPG 422 Service Manual Download Page 119](http://html1.mh-extra.com/html/tektronix/spg-422/spg-422_service-manual_1077788119.webp)
Performance Verification
SPG 422 Service Manual (B034000 and above)
4–27
21. Sync Amplitude
PAL: –300 mV
±
15 mV
a. Connect the equipment as shown in Figure 4–7.
b. Set the SPG 422 to output PAL Black Burst on Black 1 and Black 2.
c.
Display CH A on the waveform monitor.
d. Check using the waveform monitor WFM+CAL, that the sync amplitude
is –300 mV
±
15 mV.
e.
Display CH B1 on the waveform monitor.
f.
Check using the waveform monitor WFM+CAL that the sync amplitude
is –300 mV
±
15 mV.
22. Burst Amplitude
PAL: 300 mV
p-p
±
15 mV
a. Connect the equipment as shown in Figure 4–7.
b. Set the SPG 422 to output PAL Black Burst on Black 1 and Black 2.
c.
Display CH A on the waveform monitor.
d. Check using the waveform monitor WFM+CAL that the burst amplitude
is 300 mV
p-p
±
15 mV.
e.
Display CH B1 on the waveform monitor.
f.
Check using the waveform monitor WFM+CAL that the burst amplitude
is 300 mV
p-p
±
15 mV.
23. SC/H Phase Accuracy
0
°
±
5
°
a. Connect the equipment as shown in Figure 4–7.
b. Display CH A on the waveform monitor.
c.
Put the vectorscope in SCH mode.
d. Measure the SCH of the black burst using the 1781 special SCH mode.
e.
Check that the SCH is 0
°
±
5
°
.
f.
Display CH B1 on the waveform monitor.
g. Put the vectorscope in SCH mode.
h. Measure the SCH of the black burst using the 1781 special SCH mode.
i.
Check that the SCH is 0
°
±
5
°
.
Summary of Contents for SPG 422
Page 4: ......
Page 14: ...Service Safety Summary x SPG 422 Service Manual B034000 and above ...
Page 17: ......
Page 62: ......
Page 67: ......
Page 92: ......
Page 96: ...Performance Verification 4 4 SPG 422 Service Manual B034000 and above ...
Page 102: ...Performance Verification 4 10 SPG 422 Service Manual B034000 and above ...
Page 136: ...Performance Verification 4 44 SPG 422 Service Manual B034000 and above ...
Page 137: ......
Page 144: ......
Page 158: ...Maintenance 6 14 SPG 422 Service Manual B034000 and above ...
Page 159: ......
Page 162: ......
Page 223: ...9 3 SPG 422 Service Manual SPG 422 Component Digital Sync Generator FRONT PANEL 1 ...
Page 224: ...SPG 422 Service Manual 9 4 ...
Page 226: ...SPG 422 Service Manual 9 6 A2 Digital Board Static Sensitive Devices See Maintenance Section ...
Page 227: ...9 7 SPG 422 Service Manual SPG 422 Component Digital Sync Generator CPU 1 ...
Page 228: ...SPG 422 Service Manual 9 8 A7 Serial Filter ...
Page 229: ...9 9 SPG 422 Service Manual SPG 422 Component Digital Sync Generator CPU I O 2 ...
Page 230: ...SPG 422 Service Manual 9 10 ...
Page 232: ...SPG 422 Service Manual 9 12 ...
Page 234: ...SPG 422 Service Manual 9 14 ...
Page 236: ...SPG 422 Service Manual 9 16 ...
Page 238: ...SPG 422 Service Manual 9 18 ...
Page 239: ...9 19 SPG 422 Service Manual SPG 422 Component Digital Sync Generator 108 MHz OSCILLATOR 7 ...
Page 240: ...SPG 422 Service Manual 9 20 ...
Page 241: ...9 21 SPG 422 Service Manual SPG 422 Component Digital Sync Generator FINE PHASE 8 ...
Page 242: ...SPG 422 Service Manual 9 22 ...
Page 244: ...SPG 422 Service Manual 9 24 ...
Page 250: ...SPG 422 Service Manual 9 30 ...
Page 252: ...SPG 422 Service Manual 9 32 ...
Page 254: ...SPG 422 Service Manual 9 34 ...
Page 256: ...SPG 422 Service Manual 9 36 ...
Page 258: ...SPG 422 Service Manual 9 38 ...
Page 260: ...SPG 422 Service Manual 9 40 ...
Page 264: ...SPG 422 Service Manual 9 44 ...
Page 265: ...9 45 SPG 422 Service Manual 53654 48 1018 2 94 4 7 857 PART OF A5 OPTION 1 BOARD ...
Page 266: ...SPG 422 Service Manual 9 46 ...
Page 268: ...SPG 422 Service Manual 9 48 ...
Page 270: ...SPG 422 Service Manual 9 50 ...
Page 274: ...SPG 422 Service Manual 9 54 ...
Page 276: ...SPG 422 Service Manual 9 56 ...
Page 277: ......
Page 278: ......
Page 286: ...SPG 422 Service Manual 10 8 ...