![Tektronix PFG 5105 Instruction Manual Download Page 68](http://html1.mh-extra.com/html/tektronix/pfg-5105/pfg-5105_instruction-manual_1077779068.webp)
TEST?
Command T
y
pe:
Query
Query S
y
ntax:
TEST?
Query Response:
TEST < num > ;
Discussion:
This command causes the instrument to execute
internal checkout routines. A code in the res
p
onse
indicates the test results. If the instrument successfully
p
asses the test, the code number is O; a failure is
defined by the error code number returned, as follows:
o - No error detected.
1
-
An error was detected.
During the test, out
p
ut is turned off and the GPIB
p
ort remains active; out
p
ut is resumed following test
com
p
letion.
PFG 5105/5505 Instruction Manual
Command T
y
pe:
Setting or query
Setting S
y
ntax:
TRIG INT
TRIG E
X
T
TRIG MANual
Examples:
TRIG INT
TRIG E
X
T
TRIG MAN
Query S
y
ntax:
TRIG?
TRIG
Query Response (one of the following):
TRIG IN;
TRIG E
X
T;
TRIG MAN;
Discussion:
Programming
This command selects the trigger source as s
p
ecified
by the argument.
INT
Selects an internal trigger. This trigger can
not be used in arbitrary and s�ee
p
modes.
E
X
T
Selects an external trigger. The source is
the signal a
pp
lied to the TRIG IN connector.
MAN
Selects the manual trigger source. Pressing
the MAN button generates a trigger.
The query returns the current trigger source selection.
The
p
ower-on setting is TRIG MAN.
3-23
Summary of Contents for PFG 5105
Page 1: ...PFG 5105 PF G 5505 Programmable Pulse Function Generator...
Page 9: ...i I...
Page 27: ......
Page 87: ......
Page 89: ......