![Tektronix PFG 5105 Instruction Manual Download Page 40](http://html1.mh-extra.com/html/tektronix/pfg-5105/pfg-5105_instruction-manual_1077779040.webp)
Operating Instructions
GENERATING AN OUTPUT FUNCTION
The PFG 51 05/5505 provides the following output
functions: single pulse, double pulse, sine-wave, tri
angle-wave, square-wave, and de. To output a function
requires selecting the front-panel functions, param
eters, modes, etc., and their values, that define the
waveshape of the output signal. These selections are
presented in Tables 2-5 and 2-6 by general char
acteristic. The list is a guide to setting the con
trols to define the output; additional settings are
then described that add capabilities useful for special
applications.
Table 2-5
GENERATING SINE, TRIANGLE, SQUARE AND DC FUNCTIONS
Purpose
Select Output Function:
(Press one)
Select Output
Frequency:
Select Output
Amplitude:
Select Trigger Source:
(Press one)
Select Output Mode:
(Press one)
Turn Output On:
Purpose
Set to Period Input
Mode:
Set Period:
Set Delay:
Set Width:
Select Trigger Source:
Select Output Mode:
Select Pulse Output:
Turn Output On:
Step 1
SINE, TRIANGLE,
SQUARE, DC
FREQ
AMPL
I NT, MAN, EXT
CONT, TRIG, GATED,
SYNT, BURST
OUTPUT (LED on)
Step 2
If DC output is selected, enter de voltage: (DATA
numeric, units keys) ; press ENTER.
Enter frequency: (DATA numeric, units keys) ; press
ENTER.
Enter amplitude: (DATA numeric, units keys) ; press
ENTER.
If INT is selected, set internal trigger rate: (RATE,
DATA numeric, units keys), press ENTER.
If BURST is selected, set number in burst: (N BURST,
DATA numeric, units keys), press ENTER.
Table 2-6
GENERATING PULSE FUNCTIONS
Step 1
SPCL
FREQ (PERIOD
displayed)
DELAY
WIDTH
I NT, MAN, or EXT
CONT, TRIG, GATED,
SYNT, or BURST
SPULSE, DPULSE
OUTPUT (LED on)
Step 2
Enter 21
o.
Press ENTER. Change to Period with
I NCREMENT key.
Enter period: (DATA numeric, units keys) ; press
ENTER. Period is displayed instead of frequency.
Enter delay: (DATA numeric, units keys) ; press ENTER.
Delay must be less than the width.
Enter width: (DATA numeric, units keys) ; press ENTER.
Width must be greater than delay in double pulse.
If INT is selected, set trigger rate: (DATA numeric,
units keys) ; press ENTER.
If BURST is selected, set number in burst: (N BURST,
DATA numeric, units keys), press ENTER.
If error occurs, check period, delay, and width for
illegal combinations (see Width, Delay and Period
Relationships later in this section).
(NOTE: SPCL 21 O returns period mode to frequency
mode.)