![Tektronix PFG 5105 Instruction Manual Download Page 44](http://html1.mh-extra.com/html/tektronix/pfg-5105/pfg-5105_instruction-manual_1077779044.webp)
r
l
(
Double Pulse Mode
When in double pulse mode, two other factors (Equa
tions Three and Four) affect legal settings:
Equation Three:
dela
y
>
width
Equation Four:
dela
y
>
width + recover
y
time
Recover
y
time is determined b
y
pulse range as shown
in Table 2-7.
EXAMPLE
S
:
If width = 400 ns and dela
y
= 390 ns, a dela
y
�
width error will be returned.
If width = 400 ns and dela
y
= 41 O ns, then dela
y
- width 1
o
ns. This is less than the recover
y
time
for a width of 400 ns listed in Table 2-7. Therefore,
a dela
y
:,; width + recover
y
time error will be
returned.
PF G 5105/5505 Instruction Manual
Operating Instructions
Table
2-7
MINIMUM
RECOVERY TIME
W
i
d
th Range
Recovery Time
0.04 - 0.099 µs
40 ns
0.100 - 0.999
µs
50 ns
1.00 - 9.99 µs
200 ns
1 0.0 - 99.9 /lS
2.0 JlS
1 00 - 999 JtS
20 /lS
1.00 - 9.99 ms
200 JlS
10.0 - 99.9 ms
2.0 ms
Synthesizer
The s
y
nthesizer mode is available onl
y
on instruments
that include Option 02. The s
y
nthesizer outputs a
continuous waveform that is frequenc
y
-locked to an
internal quartz cr
y
stal. The displa
y
allows 4.5 digit
resolution.
2- 1 7
Summary of Contents for PFG 5105