RATE
Command Type:
Setting or query
Setting Syntax:
RATE < rate > [: < units > ]
Examples:
RATE 999.9
RATE 100:NS
Query Syntax:
RATE?
Query Response:
RATE < rate > ;
RATE 1 0E-6:S;
Discussion:
This command sets the internal trigger interval. The
power-on setting is 1 0 microseconds; the power-on
units of measure is seconds. Units can be specified
in S, mS, uS.
The rate generator is used for internal trigger and
sweep rate.
PFG 5105/5505 Instruction Manual
Command Type:
Operational
Syntax:
RECall < num >
Examples:
REC 3
RECALL 8
Discussion:
Programming
RECALL
This command changes the PFG 5105/5505 settings
to those stored in the settings buffer specified in the
argument, except for the following settings which are
not stored in settings buffers and therefore remain
unchanged:
RQS
USER
Increment Size
SPCL function 220 -Display intensity/backlight
SPCL function 240 -G PIB address
SPCL function 241 -G PIB message terminator
The argument range is 0 to 99. Buffer 0 is a read-only
buffer that contains the power-on settings. (All other
settings buffers contain the power-on settings until
they are changed by the user.)
3- 1 9