Displaying a Signal Spectrum
YBT250 Field Transmitter & Interference Tester User Manual
3- 5
Setting Trace Display Options
You can change how a trace is displayed to make it easier to analyze
a signal. Table 3--2 lists the available trace types.
Table 3- 2: Trace display options
Trace type
Description
Normal
Each point in the trace represents a single acquisition. Updates
display with each new acquisition.
Average
Each point in the trace represents an average of a number of
acquisitions. The number of acquisitions averaged is set on the
Spectrum tab of the Settings page.
Min Hold
Each point in the trace represents the minimum amplitude
measured at that frequency since the acquisition was started.
Acquisition can be restarted by tapping the restart button.
Updates displayed point only if new point < old point.
Max Hold
Each point in the trace represents the maximum amplitude
measured at that frequency since the acquisition was started.
Acquisition can be restarted by tapping the restart button.
Updates displayed point only if new point > old point.
Min/Max Hold
This is a combination of Min Hold and Max Hold. Each point in
the trace has both a minimum and maximum value that
correspond, respectively, to the minimum and maximum values
measured at that frequency since the acquisition was started.
Acquisition can be restarted by tapping the restart button.
Draws a vertical line between Max Hold and Min Hold points.
Setting the Trace Type
The default trace type is Normal. However, you can set the trace
type to any of the other four trace types.
To change the trace type, select the desired trace type from the
Trace
drop-down list. See Figure 3--2.
Summary of Contents for NetTek YBT250
Page 4: ......
Page 12: ...Table of Contents viii YBT250 Field Transmitter Interference Tester User Manual...
Page 19: ...Getting Started...
Page 20: ......
Page 39: ...Operating Basics...
Page 40: ......
Page 60: ...Operating Basics 2 20 YBT250 Field Transmitter Interference Tester User Manual...
Page 72: ...Connecting a Signal 2 32 YBT250 Field Transmitter Interference Tester User Manual...
Page 90: ...Changing Instrument Settings 2 50 YBT250 Field Transmitter Interference Tester User Manual...
Page 101: ...Reference...
Page 102: ......
Page 136: ...Displaying a Signal Spectrum 3 34 YBT250 Field Transmitter Interference Tester User Manual...
Page 170: ...Locating Interference 3 68 YBT250 Field Transmitter Interference Tester User Manual...
Page 259: ...Appendices...
Page 260: ......
Page 290: ...Appendix B Accessories B 4 YBT250 Field Transmitter Interference Tester User Manual...
Page 315: ...Glossary...
Page 316: ......
Page 322: ...Glossary Glossary 6 YBT250 Field Transmitter Interference Tester User Manual...
Page 323: ...Index...
Page 324: ......