Analyze Waveform or Trace Data
Amplitude measurements (cont.)
Measurement
Description
Positive
Overshoot
This is measured over the entire waveform or gated region and is expressed as:
Positive Overshoot = (Maximum – High) / Amplitude x 100%.
Negative
Overshoot
This is measured over the entire waveform or gated region and is expressed as:
Negative Overshoot = (Low – Minimum) / Amplitude x 100%.
Mean
The arithmetic mean over the entire waveform or gated region.
Cycle Mean
The arithmetic mean over the
fi
rst cycle in the waveform or the
fi
rst cycle in the gated
region.
RMS
The true Root Mean Square voltage over the entire waveform or gated region.
Cycle RMS
The true Root Mean Square voltage over the
fi
rst cycle in the waveform or the
fi
rst
cycle in the gated region.
Miscellaneous measurements
Measurement
Description
Positive Pulse
Count
The number of positive pulses that rise above the mid reference crossing in the waveform
or gated region.
Negative Pulse
Count
The number of negative pulses that fall below the mid reference crossing in the waveform
or gated region.
Rising Edge
Count
The number of positive transitions from the low reference value to the high reference
value in the waveform or gated region.
Falling Edge
Count
The number of negative transitions from the high reference value to the low reference
value in the waveform or gated region.
124
MDO4000 Series Oscilloscopes User Manual
Summary of Contents for MDO4054-3
Page 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Page 3: ......
Page 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Page 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Page 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...