Display Waveform or Trace Data
6.
Push
Edit Label
to label the RF and RF
versus time traces.
7.
Push
More
to choose between side menus
to compensate the RF signal path or to
con
fi
gure the RF input probe.
Trace Types
The frequency domain window supports four spectrum traces. You may turn each of these traces on and off independently.
You can display all or some of them simultaneously.
1.
Push
Spectrum Traces
from the RF Menu
to bring up the related side menu.
Spectrum
Traces
2.
Set
Normal
to
On
to display the normal
trace.
Normal
On
| Off
3.
Set
Average
to
On
to display the average
trace. Turn multipurpose knob
a
to set the
number of waveforms to include in each
average.
Average
16
On|
Off
4.
Set
Max Hold
to
On
to display the max hold
trace.
Max Hold
On|
Off
5.
Set
Min Hold
to
On
to display the min hold
trace.
Min Hold
On|
Off
The
fi
gure to the right illustrates the different
trace types.
1.
Normal trace: Each acquisition is
discarded as new data is acquired.
2.
Max hold trace: The maximum data
values are accumulated over multiple
acquisitions of the Normal trace.
3.
Min hold trace: The minimum data
values are accumulated over multiple
acquisitions of the Normal trace.
4.
Average trace: Data from the Normal
trace is averaged over multiple
acquisitions. This is true power
averaging, which occurs before the log
conversion. Each power of 2 averaging
reduces the displayed noise by 3 dB.
MDO4000 Series Oscilloscopes User Manual
111
Summary of Contents for MDO4054-3
Page 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Page 3: ......
Page 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Page 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Page 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...