Trigger Setup
Push a side-bezel button to select
sequencing the B trigger after the A as Time
or as Events.
Time
(a) 8 ns
B Events
1
Set to
Minimum
5.
Set the other Sequence Trigger parameters
in the related side- and lower-bezel menus.
B Trigger After Delay Time
The A trigger arms the instrument.
Posttrigger acquisition starts on the
fi
rst B
edge after the trigger delay time.
Trigger on B Events
The A trigger arms the instrument.
Posttrigger acquisition starts on the nth B
event.
Quick Tips
B-trigger delay time and horizontal position are independent functions. When you establish a trigger condition using
either the A trigger alone or the A and B triggers together, you can also use the horizontal position control to delay the
acquisition by an additional amount.
When using the B trigger, the A and B trigger types can only be Edge.
92
MDO4000 Series Oscilloscopes User Manual