Trigger Setup
1.
Adjust the position (delay) time by
rotating the
Horizontal Position
knob.
2.
Turn horizontal
SCALE
to acquire the
detail that you need around the position
(delay) expansion point.
The part of the record that occurs before the trigger is the pretrigger portion. The part that occurs after the trigger is the
posttrigger portion. Pretrigger data can help you troubleshoot. For example, to
fi
nd the cause of an unwanted glitch in
your test circuit, you can trigger on the glitch and make the pretrigger period large enough to capture data before the
glitch. By analyzing what happens before the glitch, you may uncover information that helps you
fi
nd the source of the
glitch. Alternatively, to see what is happening in your system as a result of the trigger event, make the posttrigger period
large enough to capture data after the trigger.
Slope and Level
The slope control determines whether the
instrument
fi
nds the trigger point on the rising
or the falling edge of a signal.
The level control determines where on that
edge the trigger point occurs.
The oscilloscope provides a long horizontal
bar or bars across the graticule to temporarily
show the trigger level.
1.
Turn the front-panel Trigger
Level
knob
to adjust the trigger level without going
to a menu.
2.
Push the front-panel Trigger
Level
knob
to quickly set the trigger level to the
midpoint of the waveform.
82
MDO4000 Series Oscilloscopes User Manual
Summary of Contents for MDO4054-3
Page 2: ...x MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual P071291300 071 2913 00...
Page 3: ......
Page 4: ...MDO4000 Series Mixed Domain Oscilloscopes ZZZ User Manual xx www tektronix com 071 2913 00...
Page 11: ...Table of Contents iv MDO4000 Series Oscilloscopes User Manual...
Page 19: ...Preface xii MDO4000 Series Oscilloscopes User Manual...