![Tektronix DSA8300 Service Manual Download Page 85](http://html1.mh-extra.com/html/tektronix/dsa8300/dsa8300_service-manual_1077994085.webp)
Removal and Installation
Time Base circuit board.
You will need a T-15 Torx driver and a 5/16 nut driver.
The Time Base circuit board is not customer-replaceable and cannot be ordered
from Tektronix. (See page 4-34,
1.
To access the Time Base circuit board, remove these assemblies:
Trim (all) (See page 4-9.)
Bottom cover (See page 4-10.)
Left and Right covers (See page 4-10.)
2.
To remove the Time Base circuit board:
(See Figure 4-16 on page 4-35.)
a.
Set the instrument with the top down on the work surface and the
instrument bottom facing you.
b.
Read all of the instructions and cautions in the
Acquisition System
section.
c.
Remove parts as necessary to access the cable or mechanical part that you
are going to replace.
(See Figure 5-6 on page 5-15.)
(See Figure 5-7 on page 5-16.)
(See Figure 5-8 on page 5-17.)
(See Figure 5-9 on page 5-18.)
3.
To reinstall the Time Base assembly:
Reverse the steps. Be sure to connect the
front-panel screws properly. (See Figure 4-17.) Refer to the cable connections
table for information about where to connect cables. (See Table 5-1.)
DSA8300 Digital Serial Analyzer and Modules Service Manual
4–37
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......