Removal and Installation
CAUTION.
Make sure you reinstall the Torx drive screw and threaded posts in the
correct position. The 3/16 inch hex threaded posts have etched lines around their
mounting holes. Interchanging the threaded posts with the Torx drive screws
could cause damage to your module connector. (See Figure 4-17.)
Do not interchange the small and large sampling module threaded posts. The large
slot sampling module threaded posts are single threaded posts and the small slot
sampling module posts are double threaded posts. Installing the double threaded
posts into the large module slot could damage the large module connector.
One of the nine T-10 Torx screws secures a 5/16 inch hex nut and grounding lug
to the inside chassis. When removing the top right T-10 Torx drive screw, make
sure this ground nut is retained for later reassembly.
Figure 4-17: T-10 screws and threaded posts