Removal and Installation
3.
Connect one end of a GPIB cable to the GPIB port of the instrument. Connect
the other end of the cable to the GPIB port of a PC.
CAUTION.
For the next step, make sure that you do not have any user interface
dialogs active. Dialogs such as Compensation, System properties will prevent the
instrument from responding to GPIB commands.
5.
Enter the following GPIB commands to program the serial number (printed
on the sampling module cover) into the replacement sampling module chassis:
a.
SYSTem:PROTect OFF
b.
SYSTem:PROPerties:CH(1-8):SERialnum <serialNum_string>
Example:
SYST:PROP:CH3:SER "B010123"
NOTE.
For a 2-channel sampling module, either channel can be speci
fi
ed in
step b. For an 80E01 sampling module, use only the lower numbered channel
number, such as CH1 or CH3.
c.
SYSTem:PROTect ON
6.
Verify the serial number change: Pull down the
Utility
menu, and then select
System Properties, Sampling Modules.
Click the channel you want to
verify, such as
C3 or C4.
This display will con
fi
rm the sampling module
serial number.
7.
Run compensation from the Utilities menu to verify that the instrument meets
its accuracy speci
fi
cations.
DSA8300 Digital Serial Analyzer and Modules Service Manual
4–47
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......