Troubleshooting
Table 4-2: Possible causes of instrument failure (cont.)
Failure symptom
Possible cause(s) / solutions
80C0X modules have no signals displayed
Module not fully installed in the slot.
Wrong channel is selected, Waveform Selector and Setup menus.
Signal cable not connected properly.
Replace optical cables.
Replace optical module or Time Base assembly.
80C0X modules; poor noise performance
Male
fi
ber-
fi
ber connections need to be cleaned.
80C0X modules; frequency response exhibits
a steep negative "droop" in its plot.
Bad internal O/E RF ampli
fi
er de-coupling. Check tightness of all rigid cable
interconnects.
BIOS error messages
Refer to the BIOS error message table. (See Table 4-3.)
Instrument locks up
Power off the instrument, and then restart.
Replace the ATX assembly.
Replace the Power PC circuit board assembly.
4–52
DSA8300 Digital Serial Analyzer and Modules Service Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......