Adjustment Procedures
Internal 10 MHz Adjust
Connect the DSA8300 to a signal generator and frequency counter as shown
in the following
fi
gure:
1.
Set the signal generator as follows
Frequency to 1 GHz
Amplitude to 1.0 V peak to peak
Output to on
2.
Set the frequency counter as follows:
Frequency to 200 kHz
Number of digits to 9
3.
Set the DSA8300 as follows:
Trigger source to TDR
Clock rate to 200 kHz
TDR Trigger mode to External (select Advance Trigger mode, and then
select External Trigger for TDR mode)
4.
Enter the following GPIB command to turn the instrument cal constant
protection off:
“SYST:PROT OFF”
5.
Set the calibration value to the default by entering the following GPIB
command:
“CALCOMP:DOUBLE "Internal10MHzRefFreq",
10e6
6.
Measure the TDR Clock Out frequency (Reading 1).
7.
Set the DSA8300 trigger mode to Internal.
8.
Measure the TDR Clock Out frequency (Reading 2).
9.
Enter the following GPIB command:
“CALComp:DOUBLE
“Internal10MHzRefFreq”,
(Reading2/Reading1 * 10M)
3–4
DSA8300 Digital Serial Analyzer and Modules Service Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......