Troubleshooting
Troubleshooting
This section contains information and procedures designed to help you isolate
faulty modules in the instrument. If an assembly must be replaced, follow the
removal instructions. (See page 4-5,
Required Equipment
You will need a digital voltmeter to check power supply voltages. (See page 4-55.)
Testing might also be required to correct some faults. In this case, you need the
test equipment listed in the
Adjustment Procedures
in this manual.
Check for Common Problems
Use the following table to quickly isolate possible failures.
Table 4-2: Possible causes of instrument failure
Failure symptom
Possible cause(s) / solutions
Mainframe will not power on
Power cord not plugged in.
Failed fuse.
Faulty power supply.
Faulty modules.
Main power switch (rear panel) not on.
Front panel power switch not on.
Monitor not connected properly.
Front panel light comes on (mainframe
powers on), but one or more fans will not
operate
Faulty fan cable.
Defective fan assembly.
Faulty power supply.
Mainframe powers on but no signal displayed
Wrong channel is selected; select correct channel from front-panel channel
selector or vertical setup menu.
Module not fully installed in the slot.
Signal cable not connected properly.
Wrong trigger input; check trigger cable connection to either direct or prescale
trigger input and make sure that the corresponding trigger source is selected in
the Trigger Setup menu.
4–50
DSA8300 Digital Serial Analyzer and Modules Service Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......