Removal and Installation
Fan Assembly
You will need a T15 Torx driver.
1.
To access the fan assembly, remove these assemblies:
Trim (all) (See page 4-9.)
Bottom cover (See page 4-10.)
Left cover (See page 4-10.)
2.
To remove the fan assembly:
(See Figure 4-15 on page 4-32.)
CAUTION.
Be careful when handling the fan assembly; the fan blades can be
damaged easily.
a.
Set the instrument with the bottom down on the work surface.
b.
Disconnect the left and right fan cables from the left and right extender
cables.
c.
Remove the two T-15 Torx drive screws that attach the fan assembly to
the top main chassis.
d.
Lift the fan assembly up and out from the chassis.
3.
To reinstall the fan assembly:
Reverse the steps. Be sure to connect the
cable marked “Left” to the cable extender matched with J391. Connect the
cable marked “Right” to the cable extender matched with J390. Refer to the
cable connections table for more information about where to connect cables.
(See Table 5-1.)
DSA8300 Digital Serial Analyzer and Modules Service Manual
4–31
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......