Removal and Installation
Figure 4-21: Sample of electrical module hardware removal
Replacing the Chassis
CAUTION.
Transport the sampling module chassis with terminations on the inputs.
Transport the sampling module chassis in the original container and store the
sampling modules on a metal or conductive surface.
When handling the exposed chassis sampling module only, do not touch any
exposed component on the circuit board. Hold the exposed module by the edges of
the circuit board or the rear or front housings.
1.
Reinstall the chassis into the cover by reversing the removal procedure. (See
page 4-45,
WARNING.
Do not install or remove any sampling modules while the instrument
is powered on. Electrical shock may occur. Always power the instrument down
before attempting to remove or insert any sampling module to avoid potential
injury from shock.
2.
Install the electrical sampling module into one of the instrument small module
compartments.
4–46
DSA8300 Digital Serial Analyzer and Modules Service Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......