![Tektronix DSA8300 Service Manual Download Page 35](http://html1.mh-extra.com/html/tektronix/dsa8300/dsa8300_service-manual_1077994035.webp)
Theory of Operation
8G FibreChannel (8.5 Gb/s), 10G FibreChannel FEC (11.317 Gb/s). These
fi
lter settings require no hardware
fi
lters.
An electrical clock recovery output signal is provided that can be routed to
the Tektronix 80A05 or 80A07 for clock recovery.
80C14 Optical Sampling
Module
The 80C14 Series Optical Sampling Module is a high-performance optical module
that supports high bandwidth telecom and datacom standards, including 16 GFC
Fibre Channel and 14.063 Gb/s In
fi
niband.
The 80C14 supports the following standards and data
fi
ltering rates: 8GFC
(8.500 Gb/s), OC-192/STM-64 (9.953 Gb/s), 10GBase-W (9.953 Gb/s),
10GBase-R (10.31 Gb/s), 40GBase-R4 (10.31 Gb/s), 100GBase-SR10
(10.31 Gb/s), 10GFC (10.52 Gb/s), ITU-T G.975 FEC (10.664 Gb/s), ITU-T
G.709 (10.709 Gb/s), 10 GbE FEC (11.1 Gb/s), 10GFC FEC (11.3 Gb/s),
Super FEC (12.5 Gb/s), 16GFC (14.025 Gb/s), 14G In
fi
niband FDR
(14.0625 Gb/s).
Clock recovery is supported with the use of the CR175A or CR286A Clock
Recovery instrument (purchased separately).
80C25GBE Optical
Sampling Module
The 80C25GBE module is designed to test long wavelength (1310 and 1550 nm)
signals. This module is intended to be used as a test and measurement tool for
high bandwidth telecommunications with its high optical bandwidth >65 GHz.
Supported standards or data
fi
ltering rates include 25.781 Gb/s 100GBase-xR4
and 27.739 Gb/s 100GBase-xR4 FEC.
65 GHz: The sampler bandwidth and response is set during the calibration
such that the combined system will have an Optical Bandwidth (-6 dB)
>65 GHz.
There is no clock recovery option available.
80A01 Trigger Prescale Preampli
fi
er Module
The 80A01 module is designed to increase the sensitivity of the prescale trigger
input of the DSA8300 to
≤
200 mV
p-p
.
The major function block of the module is a high sensitivity, high gain RF
ampli
fi
er. The input and output to this ampli
fi
er are routed to two identical SMA,
female connectors, labeled Input and Output at the module front panel.
The module receives power from the main instrument through a single connector
at the rear of the module. The power LED indicates the module is receiving power
through the interface connector.
DSA8300 Digital Serial Analyzer and Modules Service Manual
2–13
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......