Adjustment Procedures
10.
Measure the TDR Clock Out frequency. (Recheck)
11.
Verify that the 10 MHz measurement is within 2 kHz.
12.
Save the mainframe factory calibration constants by entering the following
GPIB command:
“CAL:SAV:FACTORY:MAIN”
13.
Update the mainframe calibration time/date/temperature stamp by entering
the following GPIB command:
“CAL”UPDATEINFO:MAIN”
14.
Enter the following GPIB command to turn the instrument cal constant
protection on:
SYST:PROT ON
End of Procedure