Removal and Installation
Acquisition System
The Acquisition system consists of the Time Base circuit board and bracket, the
Electrical Front-End circuit board and chassis, and the Optical Front-End circuit
board and chassis. The circuit boards are not customer replaceable.
If any of these circuit boards must be replaced, return the instrument to Tektronix
for repair. However, you can remove the assemblies temporarily, to gain access to
cables or mechanical parts that are customer-replaceable. Refer to the Replaceable
Parts list to
fi
nd part numbers of replaceable parts.
Removing a circuit board assembly to gain access to mechanical parts and then
reinstalling the same circuit board assembly should not affect the board calibration
if you are careful not to loosen or damage any parts.
CAUTION.
Use care not to loosen or damage any parts. This could disrupt
instrument operation and calibration.
Before removing the assembly, be sure that this will not affect your product
warranty.
4–34
DSA8300 Digital Serial Analyzer and Modules Service Manual
Summary of Contents for DSA8300
Page 1: ...xx DSA8300 Digital Serial Analyzer and Modules ZZZ Service Manual P077057200 077 0572 00 ...
Page 2: ......
Page 6: ......
Page 17: ...Operating Information ...
Page 18: ......
Page 21: ...Theory of Operation ...
Page 22: ......
Page 38: ...Theory of Operation 2 16 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 39: ...Adjustment Procedures ...
Page 40: ......
Page 46: ...Adjustment Procedures 3 6 DSA8300 Digital Serial Analyzer and Modules Service Manual ...
Page 47: ...Maintenance ...
Page 48: ......
Page 109: ...Replaceable Parts List ...
Page 110: ......