Installation Instructions
20
TDS6UP, TDS7UP, and CSA7UP Option CPU Upgrade Kit
4.
Reinstall the trim and carrying handle (refer to page 8). Make sure that the
interlock tabs and the standby power button are correctly positioned when
installing the trim.
5.
Reinstall the line cord (refer to page 7).
After you install the NLX board assembly, you need to reinstall the operating
software provided in this kit.
Use the following procedure to install the software:
1.
Install the new Windows 2000 Operating system CD. Follow the instructions
provided with the Windows 2000 Operating system CD.
2.
Install the TDS6000 or TDS7000 Series Product software CD. Follow the
instructions provided with the Product software CD.
3.
Install the Applications software CD. Follow the instructions provided with
the Applications software CD.
Verify Operation
This section contains instructions to check that the additional memory that you
installed is functioning.
1.
Connect the power cord to the rear panel of the instrument and then to the
mains power.
2.
Switch the power button to the On position.
3.
Push the Standby/On button to power on the instrument.
4.
After the instrument application opens, select Minimize from the File menu
to minimize the application.
5.
Right--click the My Computer icon.
6.
Select Properties from the pop-up menu.
7.
Select the Performance tab in the Properties dialog box.
8.
Check that Memory lists the correct amount of RAM.
9.
Click the TekScope button on the Windows toolbar to display the instrument
application.
This completes the verification of the NLX board assembly in your instrument.
Install the Operating
System Software