Reference
Serial Mask Testing & Serial Pattern Trigger User Manual
25
To create a new user mask that is not based on an existing mask, do the
following procedure.
Overview
Creating a new mask
Control elements and resources
Set instrument
settings
1.
Use the communications trigger features to trigger the
instrument on a signal. The instrument saves these
settings with the mask information.
See the instrument user manual for
information on displaying waveforms.
Create an empty
user mask
2.
From the button bar, touch Masks and select the
Masks tab.
3.
Touch the User Mask button.
4.
Touch the mask standard field to display the drop-down
list.
5.
Select None from the list.
6.
Touch the Copy Current Mask to User Mask button. If
you are asked if you want to overwrite the current user
mask, touch the Yes button.
7.
Touch the Edit User Mask button. The instrument
displays the Mask Edit control window.
Creating a New User Mask