To start and stop mask tests, do the following procedure.
Overview
Running a mask test
Control elements and resources
Prerequisites 1.
You must have set up the instrument to perform mask
testing as described in
Mask Test Setup
on page 12.
Start the mask
pass/fail test
2.
From the button bar, touch Masks and select the
Pass/Fail Results tab. The instrument opens the
Pass/Fail Results control window.
3.
Touch the Pass/Fail Test On/Off button to turn on mask
pass/fail testing. You can touch Reset prior to running
tests to clear the Pass/Fail Test Summary fields.
You can also use the Pass/Fail Test button in the
Pass/Fail Setup control window.
The instrument begins mask testing and displays the
test summary information in the Pass/Fail Test Summary
fields. If a mask has more than three segments, the
window displays a horizontal scroll bar below the Hits
per Segment field that lets you scroll the field to view
other segment hit data.
Stop the mask
pass/fail test
4.
Touch the Pass/Fail test button to turn off mask pass/fail
testing. Testing will also stop when the testing meets the
parameters in the Pass/Fail Setup control window.
Running a Mask Test
Summary of Contents for CSA7000 Series
Page 8: ...Preface vi Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 12: ...Getting Started 4 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 18: ...Operating Basics 10 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 48: ...Reference 40 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 62: ...Index 54 Serial Mask Testing Serial Pattern Trigger User Manual ...