Reference
16
Serial Mask Testing & Serial Pattern Trigger User Manual
Overview
Related control elements and resources
To mask test a waveform (cont.)
Set mask test
pass and fail
parameters
13.
Select the Pass/Fail Setup tab of the Masks control
window.
14.
Use the control knob, keypad, or pop-up keypad to enter
the number of waveforms to test (number of samples in
some modes), the failure threshold (the number of
waveforms that must fail to fail the test), and the delay
time (the time from when mask test starts to when the
instrument begins sampling).
15.
Use the Failure field buttons to set what the instrument
does when a mask test fails; have the instrument beep
(BEEP), send an SRQ out on the GPIB bus (SRQ), send
a trigger pulse out on the AUX OUT connector (AUX
Out), stop signal acquisition immediately (Stop Acq),
and/or print the instrument screen image to a printer
(Print).
16.
Touching the More button displays more failure
functions. Save Wfm saves the the waveform data of
the first waveform that causes the test to fail to a .wfm
file. Log Date saves time, date, and basic test
information of the first waveform that causes the test to
fail to an ASCII text (.txt) file.
Both files are saved to the location specified by the Path
button. The file name format is YYMMDD--HHMMSS,
where YY is year, MM is month, DD is day, HH is hour,
MM is minutes, and SS is seconds.
17.
Use the Completion field buttons to set what the
instrument does at the completion of a mask test.
Summary of Contents for CSA7000 Series
Page 8: ...Preface vi Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 12: ...Getting Started 4 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 18: ...Operating Basics 10 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 48: ...Reference 40 Serial Mask Testing Serial Pattern Trigger User Manual ...
Page 62: ...Index 54 Serial Mask Testing Serial Pattern Trigger User Manual ...