Overview
Related control elements and resources
To mask test a waveform (cont.)
Set test pass
and fail
parameters
(cont.)
18.
Use the Polarity buttons to set mask and waveform
polarity. Positive tests the positive waveform pulses.
Negative inverts the mask and tests the negative
waveform pulses. Both tests the first half of the tested
waveforms in positive polarity mode, then tests the
remaining waveforms in negative polarity mode.
19.
Toggle the Repeat button to On to set the instrument to
repeat (continue) mask testing on the completion of
each test.