Reference
Serial Mask Testing & Serial Pattern Trigger User Manual
27
The following procedure is an example of setting up the instrument to perform
mask testing on a DS1A signal. This example uses a DS1A signal and a
CSA7000 Instrument, but the example can easily be modified for other
communications signals and other instruments.
Overview
Creating a new mask
Control elements and resources
Install the test
hookup
1.
Connect your DS1A signal to CH 1 through suitable
cables, probes, or adapters.
2.
Press
DEFAULT SETUP
.
CSA7000 Instrument
Signal Source
Output
Set instrument
settings
3.
From the button bar, touch
Masks
and select the
Masks tab.
4.
Touch the
ANSI T1.102
button.
If not using an DS1A signal, touch the button
appropriate for the signal that you are using.
5.
Touch the mask standard field to display the drop-down
list.
6.
Select
DS1A (2.048 Mb/s)
from the list (if not using a
DS1A signal, select the standard appropriate for the
signal that you are using).
The mask is displayed, but may not be aligned with the
signal.
Mask Testing Example