Appendix G: Sequence File Text Format
G-2
AWG610 Arbitrary Waveform Generator User Manual
For example,
”SINE.WFM”, ”TRIANGLE.WFM”, ...
”GAUSSN.WFM”, ””, ...
””, ”TRIALGLE.WFM”, ...
When you do not define a file, NULL string (””) must be placed.
Repeat Count.
The <F3> is Repeat Count field.
<F3>:=<Repeat Count>|<Infinity>
<Repeat Count>::= 1 to 65536
<Infinity>::= 0
Enhanced Controls.
The <F4> to <F6> are Repeat Count, Wait Trigger, Goto
One, and Logic Jump, respectively.
<F4>::=<Wait Trigger>
<Wait Trigger>::= 1 (On) or 0 (Off)
<F5>::=<Goto One>
<Goto One>::= 1 (On) or 0 (Off)
<F6>::=<Logic Jump>
<Logic Jump>::= <Line Number> (range: 1 to 8000)
0 (No definition)
–1 (Jump to next line)
Note that the Logic Jump setting is effective depending on the jump settings
described in next paragraph.
Jump Settings
After the line descriptions, you place the jump setting descriptions as follows:
They can be omitted when you use the current settings.
TABLE_JUMP 0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0,0
LOGIC_JUMP –1,–1,–1,–1
JUMP_MODE LOGIC
JUMP_TIMING ASYNC
STROBE 0
Summary of Contents for AWG610
Page 4: ......
Page 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Page 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Page 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Page 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Page 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Page 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Page 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Page 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...