Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-43
Trigger Signal CH2
(5.35 V level)
CH1
Figure B-17: Trigger Signal (+5V check2)
8. Verify the Trigger level accuracy at –5V by following the substeps below:
a. Set the trigger level of AWG610.
Push Level (side).
Push –, 5 and ENTER keys in this order.
b. Set the offset level of generator.
Push Cursor,
%
,
&
,
,
keys as the low level of a pulse to be set
to –4.65V.
c.
Verify that no waveform is displayed on the oscilloscope.
Trigger Signal CH2
(-4.65 V level)
NOTE: In step 5b this
voltage level equals
-4.65 V.
Figure B-18: Trigger Signal (-5V check1)