The UTILITY Window
AWG610 Arbitrary Waveform Generator User Manual
3-165
The calibration must be performed in the following cases:
After a 20-minute warm up period
Prior to high precision waveform output
When the ambient temperature has changed more than +5
C or less than
–5
C from the previous calibration
Refer to the calibration and diagnostic screen to see if calibration has recently
been performed on the instrument. See Figure 3–56.
The calibration has completed when Done is displayed in the Calibration result
field. No calibration has been performed if the - - - is displayed. The factory reset
also causes the - - - to be displayed.
Figure 3-55: Calibration and diagnostic screen
. The calibration data in the memory may be lost if the instrument is
powered off while the calibrations are running.
Summary of Contents for AWG610
Page 4: ......
Page 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Page 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Page 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Page 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Page 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Page 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Page 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Page 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...