FG Mode
AWG610 Arbitrary Waveform Generator User Manual
3-225
Amplitude
The amplitude output voltage range is from 0.020 V
p-p
to 2.000 V
p-p
, in 1 mV
increments, terminated into a 50
W
load
.
Set the waveform amplitude using the
LEVEL / SCALE knob, the numeric buttons or the general purpose knob.
Offset
The offset range is from –1.000 V to +1.000 V, in 1 mV increments. Use the
VERTICAL OFFSET knob, the numeric buttons or the general purpose knob to
set the waveform offset level.
Offset is also used for setup of DC level.
Polarity
This menu sets the output waveform polarity. Pushing the Polarity menu button
toggles polarity between Normal and Inverted.
Duty
When you select Pulse waveform, Duty... side menu is added. The Duty cycle is
set from 0.1% to 99.9% using the numeric buttons or the general purpose knob.
Increment step size depends on the output frequency. Refer to Table 3–56 on
page 3–227.
Summary of Contents for AWG610
Page 4: ......
Page 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Page 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Page 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Page 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Page 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Page 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Page 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Page 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...