Appendix B: Performance Verification
AWG610 Arbitrary Waveform Generator User Manual
B-25
ÁÁ
ÁÁ
DMM
50
W
BNCĂcoaxialcable
InputĂconnector
+ĂBNCĆtoĆdualbanana adapter
+Ă50
W
BNCĂcoaxialĂcable
AWG610 Arbitrary Waveform Generator
CH1ĂoutputĂconnector
+ĂSMA(Ma)ĆBNC(Fe) adapter
+ĂBNC 50
W
Ăprecision terminator
+Ă50
W
ĂBNC coaxialcable
Figure B-9: Amplitude accuracy initial test hookup
2. Set the DMM controls as follows:
Mode . . . . . . . . . . . . . . . . . . . . . . . . . . VDC
Range . . . . . . . . . . . . . . . . . . . . . . . . . Auto
Input . . . . . . . . . . . . . . . . . . . . . . . . . . Front
3. Follow the substeps below to set the AWG610 Arbitrary Waveform
Generator controls and to select the sequence file:
a. Push UTILITY (front-panel)
System (bottom)
Factory Reset
(side)
OK (side).
b. Push SETUP (front-panel)
Run Mode (bottom)
Enhanced (side).
The AWG610 Arbitrary Waveform Generator is set to enhanced mode.
c.
Load the AMP1.SEQ file.
Refer to Loading Files on page B–10 for file loading procedures.
4. Push the RUN and CH1 output buttons.
The LEDs above the RUN button and CH1 output connector are on.
Summary of Contents for AWG610
Page 4: ......
Page 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Page 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Page 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Page 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Page 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Page 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Page 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Page 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...