FG Mode
3-224
AWG610 Arbitrary Waveform Generator User Manual
The output parameter menu selections are the same for each waveform except
Pulse and DC. Pulse has one extra side menu item (Duty), and DC has only one
side menu item (Offset).
Output
parameters
Figure 3-75: Output parameters
Frequency
The frequency is set with a 4-digit number from 1.000 Hz to 260.0 MHz using
the SAMPLE RATE / SCALE knob, the numeric buttons or the general purpose
knob.The internal cut-off filter used is determined by the waveform type and the
frequency selected. The cut-off frequencies are as follows;
Table 3-54: Output frequency and filter cutĆoff frequency
Waveform type
Output Frequency
Filter Cut-off Frequency
Sine
1.000ĂHz to 260.0ĂKHz
260.1ĂKHz to 2.600ĂMHz
2.601ĂMHz to 8.000ĂMHz
8.001ĂMHz to 15.00ĂMHz
20 MHz
50 MHz
100 MHz
200 MHz
Output parameters
Summary of Contents for AWG610
Page 4: ......
Page 24: ...Preface AWG610 Arbitrary Waveform Generator User Manual...
Page 216: ...Quick Editing AWG610 Arbitrary Waveform Generator User Manual...
Page 220: ...The Table Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 242: ...The Sequence Editor AWG610 Arbitrary Waveform Generator User Manual...
Page 288: ...The UTILITY Window AWG610 Arbitrary Waveform Generator User Manual...
Page 292: ...Capturing Waveforms AWG610 Arbitrary Waveform Generator User Manual...
Page 326: ...Programming Examples AWG610 Arbitrary Waveform Generator User Manual...
Page 338: ...File Management AWG610 Arbitrary Waveform Generator User Manual...
Page 424: ...Appendix B Performance Verification AWG610 Arbitrary Waveform Generator User Manual...
Page 440: ...Appendix E File Transfer Interface Outline AWG610 Arbitrary Waveform Generator User Manual...