Operating basics
Quick Tip
You cannot modulate or sweep noise or a DC waveform.
Generate a Burst Waveform
The instrument can output a burst using standard waveforms such as sine, square,
ramp, and pulse, or arbitrary waveforms. The instrument allows you to use the
following two types of burst modes:
Triggered Burst Mode.
A speci
fi
ed number (burst count) of waveform cycles
are output when the instrument receives a trigger input from the internal trigger
source, an external trigger source, a remote command, or the Manual Trigger
button.
Gated Burst Mode.
The instrument outputs a continuous waveform when an
effective gate signal is applied externally, when the Manual Trigger button is
depressed, when a remote command is applied, or during 50% of the selected
internal trigger interval.
To Generate a Triggered
Burst Waveform
The following example describes how to generate a double pulse using the burst
mode.
1.
Select Pulse as an
output waveform
and then push the
front-panel Burst
button.
2.
Con
fi
rm that 1-Cycle,
N-Cycles, or Inf-Cycles
is selected, which
means triggered burst
mode is enabled.
To generate double
pulse, set the burst
count (N-Cycles) to 2.
3.
This is an example of
double pulse.
4.
This waveform is a
trigger output signal.
54
AFG3000 and AFG3000C Series User Manual
Summary of Contents for AFG3000 Series
Page 2: ......
Page 3: ...AFG3000 and AFG3000C Series Arbitrary Function Generators ZZZ User Manual xx...
Page 6: ......
Page 9: ...List of Figures Figure 1 Fuse and fuse adapter 13 AFG3000 and AFG3000C Series User Manual iii...
Page 12: ...Environmental Considerations vi AFG3000 and AFG3000C Series User Manual...
Page 104: ...Operating basics 90 AFG3000 and AFG3000C Series User Manual...
Page 110: ...Application Examples 96 AFG3000 and AFG3000C Series User Manual...