Instrument interface, front panel, and rear panel
Default Setup
When you want to restore the instrument settings to the default values, use the
front-panel Default button.
1.
Push the front-panel Default
button.
2.
A con
fi
rmation pop-up message
appears on the screen.
Push OK to recall the default
settings.
Push Cancel to cancel the recall.
3.
If you select OK, the instrument
displays a 1 MHz frequency,
1 Vp-p amplitude sine waveform
as the default setup.
Quick Tips
The
AFG3000
Series
Arbitrary
Function
Generators
Programmer
Manual
describes
the
default
setup
settings
in
detail.
This
manual
is
available
on
the
accompanying
Documentation
CD.
The front-panel Default button does not reset the following settings:
Language option
Power-on settings
System related settings (display contrast, screen saver, click tone, and
beeper)
Saved setups and arbitrary waveform data
Calibration data
30
AFG3000 and AFG3000C Series User Manual