Getting started
1.
Push the front-panel Utility
button.
2.
Push the -more- bezel
button.
3.
Push the
Diagnostics/Calibration
bezel button.
4.
To execute the instrument
diagnostics, push the
Execute Diagnostics
bezel button.
To execute self calibration,
push the Execute
Calibration bezel button.
5.
If Diagnostics completes
without any errors, the
message “PASSED” is
displayed.
CAUTION.
Do not power off the instrument while executing self calibration. If the
power is turned off during self calibration, data stored in the internal memory
may be lost.
Quick Tips
Before executing self calibration, ensure that the ambient temperature is
b20 °C and +30 °C (+68 °F to +86 °F). Allow a 20 minute warm-up
period before executing self calibration.
Disconnect all the cables from the instrument when you perform self test or
self calibration.
Perform self calibration at least once a year to maintain DC accuracy. It is
recommended that the self calibration should be performed along with a
periodic check.
10
AFG3000 and AFG3000C Series User Manual
Summary of Contents for AFG3000 Series
Page 2: ......
Page 3: ...AFG3000 and AFG3000C Series Arbitrary Function Generators ZZZ User Manual xx...
Page 6: ......
Page 9: ...List of Figures Figure 1 Fuse and fuse adapter 13 AFG3000 and AFG3000C Series User Manual iii...
Page 12: ...Environmental Considerations vi AFG3000 and AFG3000C Series User Manual...
Page 104: ...Operating basics 90 AFG3000 and AFG3000C Series User Manual...
Page 110: ...Application Examples 96 AFG3000 and AFG3000C Series User Manual...