62
AFG1000 Series Arbitrary Function Generator Programmer Manual
Command Descriptions
{MINimum|MAXimum}
Arguments
<amplitude>::=<NR3>[<units>]
where:
<NR3> is the output amplitude.
<units>::=[Vpp]
Returns
<amplitude>
Examples
SOURce1:VOLTage:LEVel:IMMediate:AMPLitude 1Vpp
sets the CH 1 output amplitude to 1 Vpp.
SYSTem:ERRor[:NEXT]? (Query Only)
This query-only command returns the contents of the Error/Event queue.
Group
System
Syntax
SYSTem:ERRor[:NEXT]?
Arguments
None
Returns
<Error/event number>::=<NR1>
<Error/event description>::=<string>
Examples
SYSTEM:ERROR:NEXT?
might return the following response:
-201,"Invalid while in local"
If the instrument detects an error or an event occurs, the event number and event
message will be returned.