AFG1000 Series Arbitrary Function Generator Programmer Manual
29
Command Descriptions
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:BURSt:NCYCles
{<cycles>|INFinity|MINimum|MAXimum}
[SOURce[1|2]]:BURSt:NCYCles
? {MINimum|MAXimum}
Arguments
<cycles>::=<NRf>
where:
<NRf> is the burst count. The burst count ranges from 1 to 1,000,000.
INFinity sets the burst count to infinite count.
MINimum sets the burst count to minimum count.
MAXimum sets the burst count to maximum count.
Returns
<cycles>
Examples
SOURce1:BURSt:NCYCles 2
sets the CH 1 burst count to 2.
[SOURce[1|2]]:BURSt:SOURce
This command sets or queries the trigger source in the burst mode for the
specified channel. This command is available only in the Triggered burst mode.
Conditions
For the AFG1022, use
[SOURce1]
. Modulation, sweep, and burst conditions do
not apply to CH2 for the AFG1022.
Group
Source
Syntax
[SOURce[1|2]]:BURSt:SOURce
[TIMer|MANual|EXTernal]
[SOURce[1|2]]:BURSt:SOURce?
Arguments
TIMer
specifies an internal clock as the trigger source.
MANual
specifies a manual trigger input as the trigger source.