AFG1000 Series Arbitrary Function Generator Programmer Manual
61
Command Descriptions
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet
This command sets or queries the offset level for the specified channel.
Group
Source
Syntax
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet
{<voltage>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate]:OFFSet
?
{MINimum|MAXimum}
Arguments
<voltage>::=<NR3>[<units>]
where:
<NR3> is the offset voltage level.
<units>::=[mV | V]
Returns
<voltage>
Examples
SOURce1:VOLTage:LEVel:IMMediate:OFFSet 500mV
sets the CH 1 offset level to 500 mV.
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
This command sets or queries the output amplitude for the specified channel.
Units
Amplitude resolution
Vpp
1 mVp-p or four digits
Set the units of output amplitude by using the bezel menu selection.
Group
Source
Syntax
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
{<amplitude>|MINimum|MAXimum}
[SOURce[1|2]]:VOLTage[:LEVel][:IMMediate][:AMPLitude]
?