TDS 410, 420 & TDS 460 Service Manual (Company Confidential Final Draft)
4Ć13
Performance Tests
This subsection contains a collection of procedures for checking that the
TDS 400 Digitizing Oscilloscopes perform as warranted.
The procedures are arranged in four logical groupings:
Signal Acquisition
System Checks, Time Base System Checks, Triggering System Checks, and
O
utput Ports Checks. They check all the characteristics that are designated
as checked in Section 1,
Specification. (The characteristics that are checked
appear in boldface type under
Warranted Characteristics in Section 1.) You
can use form at the end of this section as a test record.
These procedures
extend the confidence level provided by the basic
procedures described on page 4-1. The basic procedures should be
done first, then these procedures performed if desired.
Prerequisites
The tests in this subsection comprise an extensive, valid confirmation of
performance and functionality when the following requirements are met:
H
The cabinet must be installed on the digitizing oscilloscope.
H
You must have performed and passed the procedures under
Self Tests,
found on page 4-4, and those under
Functional Tests, found on page 4-6.
H
A signal-path compensation
must have been done within the recom-
mended calibration interval and at a temperature within
_
C of the
present operating temperature. (If at the time you did the prerequisite
Self
Tests, the temperature was within the limits just stated, consider this
prerequisite met.)
H
The digitizing oscilloscope must have been last adjusted at an ambient
temperature b20
_
C and +30
_
C, must have been operating for
a warm-up period of at least 20 minutes, and must be operating at an
ambient temperature between 0
_
C and +50
_
C. (The warm-up require-
ment is usually met in the course of meeting the first prerequisite listed
above.)
Related Information
—Read
General Instructions and Conventions that
start on page 4-1. Also, if you are not familiar with operating this digitizing
oscilloscope, read
General Operating Instructions in Section 2 of this service
manual before doing any of these procedures.
Equipment Required
These procedures use external, traceable signal sources to directly check
warranted characteristics. The required equipment list follows this introduc-
tion.
Summary of Contents for 410
Page 9: ...Contents x ...
Page 13: ...Contents xx ...
Page 19: ...Safety Summary xx ...
Page 25: ...Introduction x x ...
Page 26: ...Specifications ...
Page 47: ...Operating Information ...
Page 63: ...Operating Information Operating Information 2Ć16 ...
Page 64: ...Theory of Operation ...
Page 68: ...Circuit Description Theory of Operation 3Ć4 ...
Page 69: ...Performance Verification ...
Page 81: ...Brief Procedures Performance Verification 4Ć12 ...
Page 119: ...Performance Tests Performance Verification 4Ć50 ...
Page 120: ...Adjustment Procedures ...
Page 147: ...Maintenance ...
Page 151: ...Maintenance Information Maintenance 6Ć4 ...
Page 240: ...Options ...
Page 249: ...Electrical Parts List ...
Page 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Page 252: ...Diagrams ...
Page 258: ...Diagrams 9Ć6 ...
Page 259: ...Mechanical Parts List ...
Page 263: ...Mechanical Parts List 10Ć4 ...
Page 275: ...Mechanical Parts List 10Ć16 ...