Brief Procedures
TDS 410, 420 & TDS 460 Service Manual (Company Confidential Final Draft)
4Ć11
d.
Verify the delayed trigger counter:
H
Press the main-menu button Delay by
Time.
H
Press the side-menu button Events, just below the Triggerable
after Time selection.
H
Use the General Purpose knob to enter an event count of 325
events.
H
Verify that the trigger READY indicator on the front panel flashes
about once every second as the waveform is updated on-screen.
4.
Remove the test hookup: Disconnect the standard-accessory probe from
the channel input and the probe-adjust terminal.