Warranted Characteristics
TDS 410, TDS 420 & TDS 460 Technical Reference
1Ć15
Table 1-10: Warranted Characteristics — Time Base System (Cont.)
Name
Description
Accuracy, Delta Time
Measurement
1, 2
For single-shot acquisitions using sample or high-resolution acquisi-
tion modes and a bandwidth limit setting of 100 MHz:
1 WI + 150 ppm of |Reading| + 650 ps)
For repetitive acquisitions using average acquisition mode with
≥
8 av-
erages and a bandwidth limit setting of FULL:
1 WI + 150 ppm of |Reading| + 300 ps)
1
For input signals
≥
5 divisions in amplitude and a slew rate of
≥
2.0 divisions/ns at the delta time measurement points. Signal
must have been acquired at a volts/division setting
≥
5 mV/division and not in Events mode.
2
The WI (waveform interval) is the time between the samples in the waveform record. Also, see the footnotes for
Sample Rate
Range and Equivalent Time or Interpolated Waveform Rates in Table 1-3 on page 1-8.
Table 1-11: Warranted Characteristics — Triggering System
Name
Description
Accuracy, Trigger Level or Thresh-
old, DC Coupled
Offset
1
| + 0.2 div
volts/div s Offset
Accuracy) for any channel as trigger source and for signals having
rise and fall times
≥
20 ns.
Sensitivity, Edge-Type Trigger, DC
Coupled
2
0.35 division from DC to 50 MHz, increasing to 1 division at 350 MHz
(TDS 410 and TDS 420) or 500 MHz (TDS 460) for any channel as
trigger source
Sensitivity, Video-Type, TV Field and
TV Line
2
0.6 division of video sync signal
Pulse Width, minimum, Events-Delay
5 ns
Auxiliary Trigger Input
Connector: BNC at rear panel
Input Load: equivalent to three TTL gate loads
Input Voltage (maximum): –5 VDC to +10 VDC
Auxiliary Trigger, Maximum Input
Frequency
10 MHz
Duty Cycle
High and low levels must be stable for
≥
50 ns
1
Net Offset = Offset – (Position x Volts/Div). Net Offset is the voltage level at the center of the A-D converter dynamic range. Offset
Accuracy is the accuracy of this voltage level.
2
The minimum sensitivity for obtaining a stable trigger. A stable trigger results in a uniform, regular display triggered on the selected
slope. The trigger point must not switch between opposite slopes on the waveform, and the display must not “roll” across the screen
on successive acquisitions. The TRIG’D LED stays constantly lighted when the SEC/DIV setting is 2 ms or faster but may flash when
the SEC/DIV setting is 10 ms or slower.
Summary of Contents for 410
Page 9: ...Contents x ...
Page 13: ...Contents xx ...
Page 19: ...Safety Summary xx ...
Page 25: ...Introduction x x ...
Page 26: ...Specifications ...
Page 47: ...Operating Information ...
Page 63: ...Operating Information Operating Information 2Ć16 ...
Page 64: ...Theory of Operation ...
Page 68: ...Circuit Description Theory of Operation 3Ć4 ...
Page 69: ...Performance Verification ...
Page 81: ...Brief Procedures Performance Verification 4Ć12 ...
Page 119: ...Performance Tests Performance Verification 4Ć50 ...
Page 120: ...Adjustment Procedures ...
Page 147: ...Maintenance ...
Page 151: ...Maintenance Information Maintenance 6Ć4 ...
Page 240: ...Options ...
Page 249: ...Electrical Parts List ...
Page 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Page 252: ...Diagrams ...
Page 258: ...Diagrams 9Ć6 ...
Page 259: ...Mechanical Parts List ...
Page 263: ...Mechanical Parts List 10Ć4 ...
Page 275: ...Mechanical Parts List 10Ć16 ...