Removal and Installation Procedures
Maintenance
6Ć24
c.
Next, grasp both the line cord and the retaining clamp and rotate
them 90 degrees counter-clockwise.
d.
Pull the line cord and clamp away to complete the removal.
4.
Remove line fuse: Find the fuse cap on the rear panel. (See Figure 6-6.)
Now, remove the fuse cap by turning it counter clockwise using a
flat-bladed screwdriver, and remove the line fuse.
5.
Reinstallation: Do in reverse steps 3 and 4 to reinstall the line fuse and
then the line cord.
Front Cover, Rear Cover, Cabinet, Rear EMI Gasket, and
Cabinet Handle and Feet
1.
Assemble equipment and locate modules to be removed:
a.
Have handy a screwdriver with a size T-15 Torx
R
tip (Items 1 and 2).
If removing the EMI gasket at the inside rear of the cabinet, also have
handy a pair of needle-nose pliers (Item 5).
b.
Make sure the oscilloscope front cover is installed; if it is not, install it
by snapping its edges over the trim ring.
c.
Locate the rear cover and cabinet in the locator diagram
Cabinet and
Front-Panel Mounted Modules, Figure 6-2.
2.
Orient the oscilloscope:
a.
Pull out on both of the hubs on the cabinet handle to unlock it for
positioning. While holding the hubs unlocked, rotate the handle
towards the bottom of the oscilloscope.
b.
Set the oscilloscope so its face is down with its front cover on the
work surface and its bottom facing you. Reference Figure 6-7 on
page 6-25 as you do the following steps.
3.
Remove the front feet: If removing a front foot for replacement, use a pair
of slip-jaw pliers to firmly grip the foot, then pull with a turning motion to
remove. Reverse the process to install the replacement foot. If no other
modules are being serviced, skip to the end (step 9) of this procedure.
4.
Remove rear cover: Remove the four T-15 Torx
R
screws securing the
rear cover to the oscilloscope. Lift off the rear cover. If no other modules
are being serviced, skip to the end (step 9) of this procedure.
5.
Remove the rear feet: If removing a rear foot for replacement, use your
fingers to press the foot out from inside the rear cover. Reverse process
to install the replacement foot. If no other modules are being serviced,
skip to the end (step 9) of this procedure.
6.
Remove the cabinet:
a.
Remove the T-15 Torx
R
screw at the left side of the oscilloscope.
b.
Grasp the two handle handle hubs and pull them outward as if to
rotate the handle.
Summary of Contents for 410
Page 9: ...Contents x ...
Page 13: ...Contents xx ...
Page 19: ...Safety Summary xx ...
Page 25: ...Introduction x x ...
Page 26: ...Specifications ...
Page 47: ...Operating Information ...
Page 63: ...Operating Information Operating Information 2Ć16 ...
Page 64: ...Theory of Operation ...
Page 68: ...Circuit Description Theory of Operation 3Ć4 ...
Page 69: ...Performance Verification ...
Page 81: ...Brief Procedures Performance Verification 4Ć12 ...
Page 119: ...Performance Tests Performance Verification 4Ć50 ...
Page 120: ...Adjustment Procedures ...
Page 147: ...Maintenance ...
Page 151: ...Maintenance Information Maintenance 6Ć4 ...
Page 240: ...Options ...
Page 249: ...Electrical Parts List ...
Page 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Page 252: ...Diagrams ...
Page 258: ...Diagrams 9Ć6 ...
Page 259: ...Mechanical Parts List ...
Page 263: ...Mechanical Parts List 10Ć4 ...
Page 275: ...Mechanical Parts List 10Ć16 ...