![Tektronix 410 Service Manual Download Page 106](http://html.mh-extra.com/html/tektronix/410/410_service-manual_1078890106.webp)
Performance Tests
TDS 410, 420 & TDS 460 Service Manual (Company Confidential Final Draft)
4Ć37
H
Subtract the trigger level readout from the DC calibration genera-
tor setting. The result is the trigger level accuracy.
H
CHECK that the trigger level is accurate to within
"
18 mV.
H
Press TRIGGER MENU. Press the main-menu button Slope;
then press the side-menu button for negative slope. (See icon at
left.)
H
Subtract the trigger level readout from the DC calibration genera-
tor setting. The result is the trigger level accuracy.
H
CHECK that the trigger level is accurate to within
"
18 mV.
4.
Disconnect the hookup:
a.
First set the output of the DC calibration generator to 0 volts.
b.
Then disconnect the cable from the generator output at the input
connector of CH 1.
Summary of Contents for 410
Page 9: ...Contents x ...
Page 13: ...Contents xx ...
Page 19: ...Safety Summary xx ...
Page 25: ...Introduction x x ...
Page 26: ...Specifications ...
Page 47: ...Operating Information ...
Page 63: ...Operating Information Operating Information 2Ć16 ...
Page 64: ...Theory of Operation ...
Page 68: ...Circuit Description Theory of Operation 3Ć4 ...
Page 69: ...Performance Verification ...
Page 81: ...Brief Procedures Performance Verification 4Ć12 ...
Page 119: ...Performance Tests Performance Verification 4Ć50 ...
Page 120: ...Adjustment Procedures ...
Page 147: ...Maintenance ...
Page 151: ...Maintenance Information Maintenance 6Ć4 ...
Page 240: ...Options ...
Page 249: ...Electrical Parts List ...
Page 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Page 252: ...Diagrams ...
Page 258: ...Diagrams 9Ć6 ...
Page 259: ...Mechanical Parts List ...
Page 263: ...Mechanical Parts List 10Ć4 ...
Page 275: ...Mechanical Parts List 10Ć16 ...