6Ć5
Inspection and Cleaning
Inspection and Cleaning describes how to inspect for dirt and damage on, and
how to clean the exterior and interior of, the TDS 410, TDS 420, & TDS 460
Digitizing Oscilloscopes. Inspection and cleaning are done as preventive
maintenance. Preventive maintenance, when done regularly, may prevent
oscilloscope malfunction and enhance its reliability.
Preventive maintenance consists of visually inspecting and cleaning the
oscilloscope and using general care when operating it.
How often to do maintenance depends on the severity of the environment in
which the oscilloscope is used. A proper time to perform preventive mainte-
nance is just before oscilloscope adjustment.
The cabinet helps keep dust out of the oscilloscope, and it is a major compo-
nent of the oscilloscope cooling system. The cabinet should normally be in
place when operating the oscilloscope. The front cover protects the front
panel and display from dust and damage. Install the front cover when storing
or transporting the oscilloscope.
Inspect and clean the oscilloscope as often as operating conditions require.
The collection of dirt on components inside can cause them to overheat and
breakdown. (Dirt acts as an insulating blanket, preventing efficient heat
dissipation.) Dirt also provides an electrical conduction path
that
could cause
an oscilloscope failure, especially under high-humidity conditions.
CAUTION
Avoid the use of chemical cleaning agents which might damage the
plastics used in this oscilloscope. Use only deionized water when
cleaning the menu buttons or front-panel buttons. Use a 75% iso-
propyl alcohol solution as a cleaner and rinse with deionized wa-
ter. Before using any other type of cleaner, consult your Tektronix
Service Center or representative.
Avoid the use of high pressure compressed air when cleaning dust
from the interior of this instrument. (High pressure air can cause
ESD.) Instead, use low pressure compressed air (about 9 psi).
General Care
Inspection and
Cleaning Procedures
Summary of Contents for 410
Page 9: ...Contents x ...
Page 13: ...Contents xx ...
Page 19: ...Safety Summary xx ...
Page 25: ...Introduction x x ...
Page 26: ...Specifications ...
Page 47: ...Operating Information ...
Page 63: ...Operating Information Operating Information 2Ć16 ...
Page 64: ...Theory of Operation ...
Page 68: ...Circuit Description Theory of Operation 3Ć4 ...
Page 69: ...Performance Verification ...
Page 81: ...Brief Procedures Performance Verification 4Ć12 ...
Page 119: ...Performance Tests Performance Verification 4Ć50 ...
Page 120: ...Adjustment Procedures ...
Page 147: ...Maintenance ...
Page 151: ...Maintenance Information Maintenance 6Ć4 ...
Page 240: ...Options ...
Page 249: ...Electrical Parts List ...
Page 251: ...Replaceable Electrical Parts List Electrical Parts List 8Ć2 ...
Page 252: ...Diagrams ...
Page 258: ...Diagrams 9Ć6 ...
Page 259: ...Mechanical Parts List ...
Page 263: ...Mechanical Parts List 10Ć4 ...
Page 275: ...Mechanical Parts List 10Ć16 ...