Calibration
— Typo
1L30
h. Increase the frequency of the signal generator until
the signal amplitude on the plug-in oscilloscope decreases
to 2.8 cm. Maintain a constant 4 cm signal amplitude on
the monitor
oscilloscope.
i. Check— Frequency of the signal generator must equal
or exceed 10 MHz.
NOTE
The Type 1L30 Unit must be plugged directly into
the oscilloscope vertical com partm ent fo r the high
frequency check.
j.
Remove the signal generator and the test oscilloscope
from the Video INPUT of the Type 1L30.
NOTES
%
€
Sf'
i:
k.
¥■-